Air Flow Sensor
            
        
* Information in this page is a reference that you created on the basis of information in the product catalog before the end of production, may be different from the current situation, such as goods for / supported standards options / price / features of the product. Before using, please check the compatibility and safety system.
last update: March 1, 2023
| Model | D6FZ-FGS1000 | ||
|---|---|---|---|
| Applicable fluid | Air or nitrogen (N2) | ||
| Working pressure | 0.99MPa max. | ||
| Meas-
 urement  | 
Flow *1 | Detection range | 1 to 1,000L/min (std) | 
| Resolution | 0.1L/min | ||
| Accuracy | ±2.0% of reading at 50 L/min (std) or more*2 | ||
| ±0.1%F.S. at less than 50L/min | |||
| Pressure | Detection range | 0 to 0.99MPa | |
| Accuracy | ±2%F.S. | ||
| Temper-
 ature  | 
Detection range | -10 to 60°C | |
| Accuracy | ±1.5% (absolute temperature) | ||
| Resist-
 ance to envi- ronment  | 
Operating temperature | -10 to 60°C (no condensation or icing) | |
| Operating humidity | 35 to 85%RH (no condensation or icing) | ||
| Vibration resistance | 10 to 55Hz double amplitude: 0.7mm, Acceleration: 50m/s2 in X, Y and Z
 directions (80min)  | 
||
| Shock resistance | 150 m/s2 in 6 directions (+/-X, +/-Y, and +/-Z directions), 3 times each | ||
| Pressure loss | Direct piping:10kPa max. (0.5MPa, at maximum flow)
 Using coupler (TL type by NAGAHORI INDUSTRY CO., LTD.): 10kPa max. (0.5MPa, at maximum flow)  | 
||
| Power supply voltage | 16 to 24 VDC ±10% ripple (p-p) 10% max. (Using single unit),
 24 VDC ±10% ripple (p-p) 10% max. (Using multiple units) *3  | 
||
| Power consumption | 2W max. | ||
| Measurement cycle | Approx. 62.5ms | ||
| Display | Display method | Status display by 2-color LED (illumination/blinking) | |
| Diplay value | Presence or absence of current-carrying, flow and error alarm | ||
| Output | Output
 interface  | 
Analog | Current output 4 to 20mA (2 contact)*4 Max. load resistance 270Ω max. | 
| ON/OFF | Open drain output (2 outputs)*5 24 VDC 50mA max.
 ON residual voltage 1.5V max., OFF leakage current 50μA max.  | 
||
| RS-485 | 2-wire half duplex communication, start-stop synch method
 Baud rate: 115.2kbps (fixed), Data bit length: 8 bits (fixed), stop bit length: 1 bit (fixed), parity: even (fixed), communications protocol: compatible with CompoWay/F  | 
||
| Outpus values | Momentary standard flow, Integrated standard flow, pressure, temperature, unit error output | ||
| Degree of protection | IP64 (except the case with the switch cover removed) | ||
| Wiring connection | M12 connector (8-pin) | ||
| Connection bore diameter | Rc1 (25A) bushing enables conversion to 15A and 20A | ||
| Material | Cable: PVC (Polyvinyl Chloride); Main unit: Aluminum die-cast; Display: Acrylic | ||
| Dimensions | 64(W) × 93(D) × 195(H) mm (excluding flange) | ||
| Weight (in package) | Approx. 1.2 Kg (Approx. 1.7 Kg) | ||
| Accessories | Instruction Sheet | ||
| Model | D6FZ-FGX21 | 
|---|---|
| Connectable sensor | D6FZ-FGT200 / D6FZ-FGT500 / D6FZ-FGS1000 | 
| Maximum number of
 mounted Sensors  | 
8 units*1 | 
| Display | 7-seg. 5-digit 2-step LCD display, auxiliary information indicator displays | 
| Recording interval | 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min. | 
| Displayed data | Momentary flow rate, integrated flow rate, pressure, temperature, charge/CO2
 conversion value  | 
| Recorded data | Momentary flow rate, integrated flow rate, volume flow rate, pressure, temperature | 
| Operation function | Conversion of integrated flow rate to charge/CO2 | 
| Recording mode | Continue mode*2, Ring mode*3 | 
| External output | Alarm output (Photocoupler output) *4 | 
| Communication interface | Ethernet (10BASE-T, 100BASE-TX) | 
| Internal storage device | Internal memory: Approx. 4200 data items when 1 unit is connected,
 approx. 650 data items when 8 units are connected  | 
| External storage device | SD card (measured value saving/set value saving and reading), Recommended SD card:
 HMC-SD291 (2 GB) *9 and HMC-SD491 (4 GB) (manufactured by OMRON) *5  | 
| Power supply voltage | DC input: 24 VDC±10% ripple (p-p) 10% max. | 
| Current consumption | 80 mA max. | 
| Operating temperature | Without Ethernet: -10°C to 40°C (no condensation or icing),
 with Ethernet: 0°C to 40°C (no condensation or icing)  | 
| Operating humidity | 35 to 85%RH (no condensation or icing) | 
| Storage humidity/
 temperature  | 
-15°C to 60°C, 20% to 85%RH (no condensation or icing) | 
| Insulation resistance | 20 MΩ (500 VDC) | 
| Withstand voltage | 1000 VAC, 50/60 Hz, 1 min. | 
| Vibration resistance | 10 to 150 Hz, 0.7 mm double amplitude,
 acceleration: 50 m/s2 for each in X, Y and Z directions for 80 min  | 
| Shock resistance | 150 m/s2 in 6 directions (+/-X, +/-Y, and +/-Z directions), 3 times each*6 | 
| Material | ABS | 
| Degree of protection | IP30 | 
| Mounting | Magnet mounting, screw mounting, hook | 
| Dimensions | 117.2(W) × 24.6(D) × 56.8(H) mm (Except protruding part) | 
| Weight (in package) | Approx. 150 g (Approx. 500 g) | 
| Accessories | Instruction Sheet, Startup Guide, T-branch connection cable*7, Alarm Output Connector*8 | 
last update: March 1, 2023